| Time HH:MM | Action | Open valves | Pressure _mbar.l/s_ (L/R), mbar (p2) |
| 1430 | - | M3, A9 | L/R=4×10-5 and dropping, P2=1.1×0-2 |
| 1437 | spray around valve 15 in GHC | M3, A9 | No change. |
| - | Open 5 | M3, A9,5 | L/R= drops to 1×10-6 |
| - | Close 5 | M3, A9 | L/R= 9.5×10-7 and dropping |
| 1448 | Open 16 | M3, A9,16 | L/R=3×10-4 and dropping (saturated at 4×10-5 in 5 mins) |
| 1448 | Close 16 and open 5 | M3, A9,5 | L/R= 2×10-6, P2=9.2×10-3 |
| 1450 | Spraying around the traps | M3, A9,5 | L/R= 1.2×10-6, P2=9.2×10-3 |
| 1450 | Spraying around GHC connections | M3, A9,5 | L/R= 1.0×10-6, P2=1.0×10-2 |
| 1452 | Close 5 | M3, A9 | - |
| 1454 | - | M3, A9 | L/R= 9.0×10-7, P2=5.0×10-3 |
| 1454 | Open 16 | M3, A9,16 | P2 drops and L/R increases; L/R= 1.5×10-4, P2=4.8×10-3 |
| 1509 | - | M3, A9,16 | L/R=3.7×10-5P2=3×10-3 |
| 1509 | Spray GHS, still lines and connections in/on GHS and Fridge and traps. | M3, A9,16 | No change; L/R=3.7×10-5P2=3×10-3 |
| 1515 | Close 16, close A9 | M3 | L/R=3.8×10-8, P2=2.4×10-3 |
| 1515 | Open A9 | M3, A9 | L/R=1.5×10-5 and dropping (saturated at 8.9×10-7 in 5 mins) |
| 1528 | Close A9 | M3 | - |
| 1535 | Attatching Speedivalve (open) to auxillary trap 'from fridge' valve in GHC (CHECK NAME) | M3, speedivalve | - |
| 1535 | Open A9 and 16, pumping with leak detector | M3, speedivalve, A9, 16 | L/R=2×10-5, P2=5×10-3 |
| 15-- | Shut 16 | M3, A9, speedivalve | L/R=2×10-6, P2=5×10-3 |
| 1554 | Open 16 | M3, A9, speedivalve, 16 | L/R=3×10-5, P2=8×10-3 |
| 1556 | - | M3, A9, 16 | L/R=1.5×10-5, P2=3.8×10-3 |
| 1557 | Close speedivalve | M3, A9, 16 | No change; L/R=3×10-5, P2=8×10-3 |
| 1558 | Open speedivalve | M3, A9, speedivalve, 16 | No change; L/R=3×10-5, P2=8×10-3 |
| 1558 | Close speedivalve | M3, A9, 16 | - |
| 1558 | Close 16 | M3, A9 | - |
| 1600 | Open 16 | M3, A9, 16 | - |
| 1614 | - | - | L/R=8.07×10-7P2=2.3×10-3 |
| 1614 | Open speedivalve | M3, A9, speedivalve, 16 | - |
| 1616 | - | M3, A9, speedivalve, 16 | L/R=1.6×10-4 |
| 1616 | Close speedivalve | M3, A9,16 | - |
| 1634 | Close A9, 16 and M3. Turn off leak detector | - | - |
| 1635 | Vent leak detector to air and allow turbo to slow. | - | - |
| Time HH:MM | Action | Open valves | Pressure _mbar.l/s_ (L/R), mbar (p2) |
| 1151 | Attach leak detector and open M3 and A9 | M3, A9 | L/R=5×10-6, P2=3.7×10-2 |
| 1200 | Open 8, 2, 0 and leave leak detector to pump down | M3, A9, 8, 2, 0 | L/R=2×10-2, P2=6.5×10-2 |
| 1225 | - | M3, A9, 8, 2, 0 | L/R=7.4×10-6, P2=9×10-3 |
| 1225 | Spray around still line, near the cryostat | M3, A9, 8, 2, 0 | L/R=7.1×10-6 |
| 122- | Spray T-bar on GHS | M3, A9, 8, 2, 0 | L/R=5.8×10-6, P2=8.3×10-3 |
| 122- | Spray back of the turbo and connections | M3, A9, 8, 2, 0 | L/R=5.0×10-6, P2=7.9×10-3 |
| 122- | Spray black vacuum manifold | M3, A9, 8, 2, 0 | L/R=9.2×10-6, P2=7.5×10-3 |
| 1400 | note: clamps lose on T-bar on GHS (clamp closest to front display) | M3, A9, 8, 2, 0 | L/R=1.1×10-6, P2=3.3×10-3 |
| 14-- | Sprayed T-bar on GHS again | M3, A9, 8, 2, 0 | L/R=1.1×10-6, P2=3.3×10-3 |
| 14-- | Sprayed still pumping line of fridge again | M3, A9, 8, 2, 0 | L/R=1.1×10-6, P2=3.3×10-3 |
| 14-- | Sparyed every connection in GHC | M3, A9, 8, 2, 0 | L/R=1.1×10-6, P2=3.3×10-3 |
| 14-- | Closed 0, 2, 8 | M3, A9 | L/R=1.1×10-6, P2=3.3×10-3 |
| 14-- | Open 5 (quick testing of traps) | M3, A9, 5 | L/R increases a littl |
| 14-- | Open 16 (quick testing of traps) | M3, A9, 5, 16 | L/R increases massively L/R=7×10-3, P2=9.5×10- |
| Time | What | P2 ( mbar) |
Pstill ( mbar) |
L/R ( mbar.l/s) |
|---|---|---|---|---|
| 1/2/2011: 1405 |
Tubo on, Open M3, A9, 8, 2, 0, 1, 6, 7, compressor bipass, gave valve | - | - | - |
| 1445 | Pturbo=4.5×10-4 mbar; Pstill=1.9×10-1 mbar. Run S1 ot pump faster (open S2) |
- | - | - |
| 1523 | Pturbo=2.8×10-4 mbar; Pstill=5×10-4 mbar | - | - | - |
| 2/2/2011: - | Pturbo=3.1×10-4 mbar; Pstill<5×10-5 mbar. Close S1 and S2 | - | - | - |
| 1130 | close M3, A9, 0, 1, 2, 8 and stop turbo. | - | - | - |
| 1134 | start leak detector | |||
| 1141 | - | 4.7×10-3 | <5×10-5 | 5×10-10 |
| 1143 | spray valves on top of leak detector (LD) and around M3 connection | 3.8×10-3 | <5×10-5 | 2.6×10-10 |
| 1143 | open M3 | - | - | - |
| 1148 | 2.8×10-3 | 6.5×10-4 | 7.4×10-10 | |
| 1150 | spary around M3 connections | 2.6×10-3 | 7.4×10-4 | 6.6×10-10 |
| 1150 |
open A9 | - |
- | - |
| 1232 | - | 1.9×10-3 | 1.9×10-3 | 2.5×10-7 |
| 1239 | spray around the back of of the gas cabinet | 1.8×10-3 | 2.1×10-3 | 2.5×10-7 |
| 1239 | open 8 | - | - | - |
| 1323 | - | 1.6×10-3 | 3.1×10-3 | 3.7/3.8×10-7 |
| 1340 | - | 1.6×10-3 | 3.4×10-3 | 3.8×10-7 |
| 1343 | spray around back of gas cabinet | 1.6×10-3 | 3.5×10-3 | 3.8×10-7 |
| 1344 | open 2 | 1.6×10-3 | 3.5×10-3 | 3.8×10-7 |
| 1359 | open 0 | 1.6×10-3 | 3.5×10-3 | 3.8×10-7 |
| 1400 | open 1 | - | - | - |
| 1403 | spray around still (on DR and gas cabinet), spray in gas cabinet. | 1.8×10-3 | 3.9×10-3 | 3.8×10-7 |
| 1404 | leak detector off, open gate valve, leak detector on | - | - | - |
| 1455 | - | 1.6×10-3 | 8.5×10-3 | 9.9×10-7 |
| 1500 | spray around still (on DR and gas cabinet), spray in gas cabinet. | 1.6×10-3 | 8.5×10-3 | 9.8×10-7 |
| 1504 | repeat the above | - | - | - |
| 1505 | - | 1.6×10-3 | 8.5×10-3 | 9.9×10-3 |
| 1508 | - | 1.6×10-3 | 8.5×10-3 | 9.8×10-3 |
| 1508 | - | 1.6×10-3 | 8.5×10-3 | 9.8×10-7 |
| 1519 | - | 1.6×10-3 | 8.4×10-3 | 9.7×10-7 |
| 1543 | spray around still (on DR and gas cabinet), spray in gas cabinet. | 1.6×10-3 | 8.5×10-3 | 9.9×10-7 |
| 1648 | 1.6×10-3 | 9×10-3 | 9.9×10-7 |
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.