3. Menu.vi will appear. Run menu.vi by clicking the white arrow, below 'Edit'. Notice the 'grid' background disappears and becomes a solid green colour. This indicates LabView is running.
| Image | What | Comments |
|---|---|---|
|
Run button | Click this to run the LabView VI |
|
Start PSU interface | This opens the Vector Magnet Controller which interfaces with the magnet Power Supply Units (PSU) |
|
Start temperature scanner | This starts the temperature scanner. Always run this first! |
|
Start sequences window | - |
|
This is not available on our system. |
5. Select '2890 Initialisation.ini' file by clicking on the folder icon and selecting the 2890 initialisation.ini file ( check that the path is correct - it MUST be C:\Cryogenic Software\DAQCard-Driven\2890 Initialisation.ini). Click 'CONFIRM'
6. 'menu.vi' will now appear. Start the temperature scanner by clicking on 'SCANNER'.
7. The Thermometer Scanner will now start. Press 'Initialise' to start. Press 'New log file' to create a new log file.
| Image | What | Comments |
|
Indicates if scanner is on | The green button is as shown when the scanner is on, and as shown below when off. |
|
Log file information | The green button flashes when the system is logging temperatures. |
|
To quit the vi | |
|
To initialise the scanner | |
|
||
|
To display the log file as a graph | |
|
To create a new log file | The log has to be started manually, start the file name with the date, followed by any other names. |
9. The Vector Magnet Control.vi window will appear. Upon starting, the software always registers a non-zero magnetic field, therefore ramp to zero first and then set the field you want. Notice: The field is non-zero (even if the magnet is at zero field to start with):
The system is reporting it is out of range:
| Image | What | Comments |
|
Coordinate system and field targets (input) | The grey button indicates which coordinate system is in use, click on it to change coordinate systems. Beneath this is where the desired target field is entered. Either use the 'up and down' click buttons, or simply enter the field in the black boxes. |
|
Field values and current ramp rate (output) | The ramp rate is not explicitly input, instead it is determined by the input field and the time to reach the desired field (rate=field/time). Therefore always manually calculate the rate in mT/s for each X, Y and Z input value. Be aware if you enter different target values for each magnet, they may have different ramp rates. |
|
Ramp path (calculated) | These X, Y & Z Path graphs show the ramp path the magnet will take. In the Vector magnet control.vi screen shot above, the magnet will ramp in the positive x-direction, positive y-direction and in the negative z-direction. These graphs are green when the system can ramp at a certain rate, and are red when they cannot ramp at the given input rate. |
|
Information on setting the field | Here you can see if the field is out of range (too high or too fast): green means the range is fine and red means it is out of range. If values are input into the field target, clicking 'calculate time' usually gives the shortest time to the input value. It is a useful lower range. Limit target..... |
|
Confirmation of field inputs | These display the actual target values in the current coordinate system, in addition to the ramp rates in A/s. This box confims your inputs only. |
|
Field status (output) | This also reports the field status, in both cartesian and spherical coordinates, as well as the Amps put into the magnets. The green lights, when lit, indicate the heaters are on. |
|
Comment text box and status bar (below) | Above the text box displays the magnet status, for all mangets. Below the status bar indicates the progress during ramping. |
|
Field ramping controls | 'Ready' indicates if the field is ready to be ramped, 'GO' starts ramping and 'ABORT' is to abort mid-ramp. 'GO TO PERSISTENT MODE' is obviously to put the magnet in persistent mode! |
|
magnet output display (not shown in full here) | Displayed here is the magnet output. When the system is first started only 'field' is shown in purple - this is because the software registers a field upon startup, even if there is not field. |
|
||
|
11. Click 'Calculate Times' so the software registers the input. Notice:
12. Select the desired time to ramp (for 'real' fields), in the case of initially "ramping" the field to zero so the software registers zero field, you can leave it at 1 second.
13. The X, Y and Z path displays will change/adjust automatically once you have finished entering the time (or fields). If they are green: start. If they are red input another time.
14. Click 'GO' to start ramping the field.
15. The text box on the vector magnet controler will read "Wating for heaters to warm up..." indicating the magnets will start soon.
16. The vector magnet controller will look as shown below. Notice:
17. When the magnet has finished ramping the text box will display:
The 'Magnet Output' will display the field (not shown as it's at zero)and path
2. You will see the text box below after aborting the field.



3. For the Z-magnet, set the VISA resource name to COM3 in the front panel. The x-magnet is COM7, and y is COM4 (see and y-magnets are on different com ports - see the "2890 Initialisation.ini" file for details.
4. make sure the 'read' and 'write' arrows are ON and glowing green. Type in the desired switch heater voltage in the "string to write" section following the syntax "SET HEATER n" and run. For example, "SET HEATER 3.0" will set the switch heater voltage to 3.0V. To check the voltage is set correctly, type in "SET HEATER" and run. The current switch heater voltage will be printed out in the


"read string" section.
You can also turn the heaters On and off by typing 'HEATER ON' and 'HEATER OFF' (while you are at zero field)
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.